Methodology of thermal drift measurements for surface guided radiation therapy systems and clinical impact assessment illustrated on the C-Rad Catalyst+ HD system

Download (2.79 MB)
journal contribution
posted on 2025-05-11, 19:27 authored by Joerg LehmannJoerg Lehmann, Therese S. Standen, Guneet Kaur, Joshua Wolf, Alex Wilfert, John SimpsonJohn Simpson
Thermal drift of optical systems employed for surface guided radiation therapy (SGRT) adds uncertainty to patient setup and monitoring. This work describes methods to measure the drift of individual camera pods as well as the drift of the combined clinical signal. It presents results for four clinical C-Rad Catalyst+ HD systems. Based on the measured clinical drift, recipes are provided on how to calculate relevant uncertainties in patient setup and patient position monitoring with SGRT. Strategies to reduce the impact of drift are explained. While the results are specific to the systems investigated, the methodology is transferable and the clinical recipes are universally applicable.

History

Journal title

Technical Innovations and Patient Support in Radiation Oncology

Volume

21

Issue

March 2022

Pagination

58-63

Publisher

Elsevier

Language

  • en, English

College/Research Centre

College of Engineering, Science and Environment

School

School of Information and Physical Sciences

Rights statement

© 2022 The Authors. Published by Elsevier B.V. on behalf of European Society for Radiotherapy & Oncology. This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/.