MR simulation for prostate radiation therapy: effect of coil mounting position on image quality

Download (1002.54 kB)
journal contribution
posted on 2025-05-11, 10:02 authored by J. Sun, P. Pichler, J. Dowling, Frederick MenkFrederick Menk, Peter StanwellPeter Stanwell, J. Arm, Peter GreerPeter Greer
Objective: To eliminate the effects of body deformation for MR-based prostate treatment planning, coil mounts are essential. In this study, we evaluated the effect of the coil set-up on image quality. Methods: A custom-designed pelvic-shaped phantom was scanned by systematically increasing the anterior body-to-coil (BTC) distance from 30 to 90 mm. The image quality near the organs of interest was determined in order to characterize the relationship between image quality and BTC distance at the critical organ structures. The half intensity reduction (HIR) was calculated to determine the sensitivity of each organ structure to the BTC distance change. Results: As the BTC distance increased, the uniformity reduced at 3% per millimetre. The HIR value indicated that the bladder signal is most sensitive to the change in BTC distance. By maintaining a constant BTC distance set-up, the intensity uniformity was improved by 28% along the B0 directions. Conclusion: Positioning the MRI coil on mounts can reduce body deformation but adversely degrades the image quality. The magnitude of this effect has been quantified for prostate MR simulation scanning. The coil needs to be positioned not only with a minimal but also uniform BTC distance in order to maximize image quality. Advances in knowledge: A method to characterize the effect on image quality due to the use of coil mounts was demonstrated. Coil mounts whose height can be adjusted individually to keep BTC distance constant are necessary to maintain a uniform image across the entire field of view.

History

Journal title

British Journal of Radiology

Volume

87

Issue

1042

Publisher

British Institute of Radiology

Language

  • en, English

College/Research Centre

Faculty of Science and Information Technology

School

School of Mathematical and Physical Sciences

Usage metrics

    Publications

    Exports

    RefWorks
    BibTeX
    Ref. manager
    Endnote
    DataCite
    NLM
    DC