Cine (continuous) mode images obtained during VMAT delivery are heavily degraded by banding artefacts. We have developed a method to reconstruct the pulse sequence (and hence dose deposited) from open field images. For clinical VMAT fields we have devised a frame averaging strategy that greatly improves image quality and dosimetric information for three-dimensional dose reconstruction.
History
Source title
Proccedings of the 7th International Conference on 3D Radiation Dosimetry (IC3DDose)
Name of conference
7th International Conference on 3D Radiation Dosimetry (IC3DDose)